Absolute consistency
Unique, Patented Piston Drive Mechanism
The new “rack rail” mechanism guarantees
identical performance on each channel. The two cogwheels running
on each side of the piston housing forces all the pistons to move
exactly the same distance.
The pressure is applied evenly and simultaneously on each piston,
resolving the common downfall of multichannel pipettes: volume disparities
between channels.
|
| |
Wide Useful Volume Range
Wide useful volume range: 10-100%
of the nominal volume.
Specifications provided at 10%, 50% and 100% of the nominal volume,
as recommended by the ISO8655 standard.
As for all Gilson pipettes, Pipetman Neo specifications are more
stringent than those required for ISO8655 compliance. On average
two times better than recommendation.
|
| |
|
|
| |
Pipetman Neo Multi
Models |
Gilson Maximum
Permissible Errors |
ISO 8655 Maximum
Permissible Errors |
| |
Model |
Diamond
Tips* |
Reference
Number |
Volume
(µl) |
Systematic
error (µl) |
Random
error (µl) |
Systematic
error (µl) |
Random
error (µl) |
| |
8x20 |
D200 or
DF30 |
F14401 |
2
10
20 |
± 0.10
± 0.20
± 0.40 |
< 0.08
< 0.10
< 0.20 |
± 0.40
± 0.40
± 0.40 |
< 0.20
< 0.20
< 0.20 |
| |
12x20 |
D200 or
DF30 |
F14402 |
2
10
20 |
± 0.10
± 0.20
± 0.40 |
< 0.08
< 0.10
< 0.20 |
± 0.40
± 0.40
± 0.40 |
< 0.20
< 0.20
< 0.20 |
| |
8x200 |
D200 or
DF200 |
F14403 |
20
100
200 |
± 0.50
± 1.00
± 2.00 |
< 0.25
< 0.40
< 0.60 |
± 3.20
± 3.20
± 3.20 |
< 1.20
< 1.20
< 1.20 |
| |
12x200 |
D200 or
DF200 |
F14404 |
20
100
200 |
± 0.50
± 1.00
± 2.00 |
< 0.25
< 0.40
< 0.60 |
± 3.20
± 3.20
± 3.20 |
< 1.20
< 1.20
< 1.20 |
* Recommended tip for establishing maximum permissible
errors (ISO 8655 requirement).
A large range of suitable Gilson Diamond tips is available.
Read more... |
|
|