Pipetman Neo is an evolution of the classic Pipetman P
that addresses the growing susceptibility of pipette
users to repetitive strain injuries.
Pipetman Neo exhibits a distinct enhancement in comfort,
a wide useful volume-range with the same legendary
robustness, accuracy and precision as Pipetman P.
|
| |

|
Up to 50% reduction of pipetting forces

Newly engineered springs provide a significant reduction of pipetting
and purge forces.
Up to 50% reduction of pipetting and purge forces on certain models.
Pipetman Neo was elaborated from the Pipetman P, maintaining its legendary
robustness, accuracy and precision.
Color-coded push-buttons
Color-coded
push-buttons facilitate easy identification of the different models
and the corresponding quality Diamond tips.
Helps differentiate Pipetman Neo pipettes from the classic Pipetman
P models.
Wide Useful Volume Range
Wide useful volume range: 10-100% of the nominal volume.
Specifications provided at 10%, 50% and 100% of the nominal volume,
as recommended by the ISO8655 standard.
As for all Gilson pipettes, Pipetman Neo specifications are more
stringent than those required for ISO8655 compliance.
|
| |
|
|
| |
Pipetman Neo Models |
Gilson Maximum
Permissible Errors |
ISO 8655 Maximum
Permissible Errors |
| |
Model |
Diamond
Tips* |
Reference
Number |
Volume
(µl) |
Systematic
error (µl) |
Random
error (µl) |
Systematic
error (µl) |
Random
error (µl) |
| |
P2N |
D10
DL10 |
F144561 |
0.2
0.5
2 |
± 0.024
± 0.025
± 0.030 |
< 0.012
< 0.012
< 0.014 |
± 0.08
± 0.08
± 0.08 |
< 0.04
< 0.04
< 0.04 |
| |
P10N |
D10
DL10 |
F144562 |
1
5
10 |
± 0.025
± 0.075
± 0.100 |
< 0.012
< 0.030
< 0.040 |
± 0.12
± 0.12
± 0.12 |
< 0.08
< 0.08
< 0.08 |
| |
P20N |
D200 |
F144563 |
2
10
20 |
± 0.10
± 0.10
± 0.20 |
< 0.03
< 0.05
< 0.06 |
± 0.20
± 0.20
± 0.20 |
< 0.10
< 0.10
< 0.10 |
| |
P100N |
D200 |
F144564 |
10
50
100 |
± 0.35
± 0.40
± 0.80 |
< 0.10
< 0.12
< 0.15 |
± 0.80
± 0.80
± 0.80 |
< 0.30
< 0.30
< 0.30 |
| |
P200N |
D200 |
F144565 |
20
100
200 |
± 0.50
± 0.80
± 1.60 |
< 0.20
< 0.25
< 0.30 |
± 1.60
± 1.60
± 1.60 |
< 0.60
< 0.60
< 0.60 |
| |
P1000N |
D1000 |
F144566 |
100
500
1000 |
± 3
± 4
± 8 |
< 0.6
< 1
< 1.5 |
± 8
± 8
± 8 |
< 3
< 3
< 3 |
* Recommended tip for establishing maximum permissible
errors (ISO 8655 requirement).
A large range of suitable Gilson Diamond tips is available.
Read more... |
|
|